
Vacuum System
Vacuum Gauges
Why measure vacuum?
A vacuum system needs instruments capable of determining whether the required pressure has been reached and maintained.
No single gauge works optimally across the entire pressure range encountered in a mass spectrometer.
Different gauges are therefore used for different pressure regimes.
Pirani gauge
A Pirani gauge measures pressure indirectly through the thermal conductivity of the gas.
A heated element loses heat to surrounding gas molecules.
As pressure changes, heat loss changes.
This allows pressure to be inferred.
Pirani gauges are particularly useful in rough and medium vacuum regions.
Ionisation gauges
At much lower pressures, ionisation-based gauges can be used.
Gas molecules are ionised and the resulting ion current provides information about pressure.
Lower gas density produces a lower ion current.
Gauge readings depend on gas
Many vacuum gauges are calibrated using a particular gas.
Consequently, the displayed pressure may not represent the exact absolute pressure when the residual gas composition differs significantly from the calibration gas.
Gauges as diagnostic tools
The trend can be as important as the absolute value.
For example:
Normal: pressure decreases steadily during pump-down.
Leak: pressure may plateau above the expected value.
Sudden gas load: pressure rises rapidly.
Outgassing: pressure may improve slowly over an extended period.
Key takeaway
Vacuum gauges convert physical properties of the residual gas into pressure measurements that allow the instrument to monitor and protect its vacuum system.


