Gas Flow
Flow Regimes
Gas does not always behave the same way. The way it moves depends on how often molecules collide with each other versus how often they hit the walls of the chamber or tube. This is described by the Knudsen number:
Kn = λ / L
where λ is the mean free path and L is a characteristic dimension (for example, the diameter of a tube or the size of the chamber).
Approximate regimes:
Continuum (viscous) flow — Kn ≪ 0.01. Molecules collide with each other far more often than with walls. Ordinary fluid behaviour applies.
Transitional (Knudsen) flow — Kn roughly 0.01 to 10. Both molecule–molecule and molecule–wall collisions matter.
Molecular (free molecular) flow — Kn ≫ 1. Molecules travel in straight lines and hit walls much more often than they hit each other.
Where These Regimes Appear in a Mass Spectrometer
Sample inlets, atmospheric-pressure sources and higher-pressure regions often operate in continuum or transitional flow.
High-vacuum analyser regions, ion guides and most of the flight path operate in molecular flow.
The transition between regimes occurs in intermediate vacuum stages (differential pumping regions).
Conductance
Conductance is a measure of how easily gas flows through a tube, aperture or orifice. In molecular flow, conductance depends mainly on geometry (diameter and length) and is independent of pressure. In continuum flow it depends strongly on pressure. This difference is important when designing or troubleshooting vacuum systems.
Key Takeaways
The Knudsen number tells you which flow regime applies.
Most of the ion path inside a mass spectrometer is in molecular flow.
Conductance behaves differently in continuum versus molecular flow.
Knowing the regime helps predict how gas and ions move through the instrument.