Common Terminology

Core Concepts

Common Terminology

Common Terminology


Core Terms

m/z (mass-to-charge ratio)

The quantity measured by a mass spectrometer. It is the mass of an ion divided by its charge number. Written in italic as m/z. For a singly charged ion, the numerical value of m/z is approximately equal to the ion’s mass in daltons.


Ion, Molecular Ion, Fragment Ion, Adduct, Charge State

  • Ion — Any atom or molecule that carries a net electrical charge.

  • Molecular ion — The ion formed from the intact molecule (often written M⁺• or [M+H]⁺).

  • Fragment ion — An ion produced when a larger ion breaks apart.

  • Adduct — An ion formed by the addition of another species (for example [M+Na]⁺ or [M+H]⁺).

  • Charge state — The number of charges on an ion (1+, 2+, 3+, etc.). Higher charge states appear at lower m/z.


Mass Spectrum and Related Mass Terms

  • Mass spectrum — Plot of ion intensity versus m/z.

  • Base peak — The most intense peak in the spectrum, often set to 100 % relative abundance.

  • Molecular weight — General term for the mass of a molecule; context determines whether monoisotopic or average is meant.

  • Monoisotopic mass — Mass calculated using the most abundant isotope of each element.

  • Average mass — Mass calculated using the average atomic weights of the elements (weighted by natural abundance).

  • Exact mass — The calculated mass of an ion with a specified isotopic composition.

  • Accurate mass — The experimentally measured mass. Accurate mass and exact mass are not the same thing.

Common pitfall: People often say “accurate mass” when they mean the calculated exact mass, or vice versa. Accurate mass is measured; exact mass is calculated.


Resolution vs Resolving Power

Both terms describe the ability to separate close m/z values. In everyday use they are often treated as interchangeable. Strictly, resolving power is usually m/Δm (a number), while resolution can refer to the peak width itself. Always check how a particular instrument or document defines the term (FWHM or valley definition).


Mass Accuracy

How close a measured m/z is to the true value.

  • Expressed in parts per million (ppm) or millidaltons (mDa).

  • Example: a 5 ppm error at m/z 500 is 0.0025 Da (2.5 mDa).


Sensitivity, Limit of Detection, Signal-to-Noise

  • Sensitivity — How little analyte can be detected; influenced by ionisation, transmission and noise.

  • Limit of detection (LOD) — The lowest amount that can be reliably distinguished from background.

  • Signal-to-noise (S/N) — Ratio of peak height to the noise level. A practical measure of whether a peak is real.


Acquisition Modes

  • Scan — The analyser sweeps a range of m/z values to produce a full spectrum.

  • SIM (Selected Ion Monitoring) — Monitors only specific m/z values.

  • SRM / MRM (Selected / Multiple Reaction Monitoring) — Targeted MS/MS modes that monitor specific precursor-to-product transitions.

  • MS/MS or tandem MS — Selection of a precursor ion, fragmentation, and analysis of the product ions.

  • Precursor ion — The ion selected for fragmentation.

  • Product ion — An ion formed from the fragmentation of the precursor.


Hard vs Soft Ionisation

  • Hard ionisation — Deposits enough energy to cause extensive fragmentation (classic example: electron ionisation).

  • Soft ionisation — Produces mostly intact molecular ions with little fragmentation (examples: electrospray, MALDI).


Unit Mass Resolution vs High Resolution

  • Unit mass resolution — Peaks separated by 1 Da are resolved (typical of many quadrupoles).

  • High resolution — Resolving power high enough to separate ions that differ by much less than 1 Da (typical of TOF, Orbitrap, FT-ICR).


Calibration, Tuning, Lock Mass

  • Calibration — Adjustment of the m/z scale using ions of known mass.

  • Tuning — Optimisation of voltages and parameters for peak shape, resolution and sensitivity.

  • Lock mass — A reference ion of known m/z introduced during acquisition to correct mass drift in real time.


Vacuum Terms That Appear Early

  • Base pressure — The pressure achieved with no sample or gas load.

  • Operating pressure — The pressure while the instrument is running with sample or reagent gas.


Other Frequently Used Terms

  • Ion current — The electrical current produced by ions arriving at the detector.

  • Transmission — The fraction of ions that successfully move from one region of the instrument to the next.

  • Space charge — Mutual repulsion between ions when their density i