Mass Spectrometry Principles
Quantitation
From Signal to Amount
Quantitation relates the intensity of an ion signal to the amount of substance present. In principle the ion current is proportional to concentration, but many factors affect the relationship.
Common Approaches
External calibration (standard curve)
Internal standards (especially stable-isotope-labelled standards)
Standard addition
Response factors
Isotope-labelled internal standards are widely used because they experience nearly identical ionisation and transmission losses as the analyte.
Factors That Affect Accuracy
Matrix effects (ion suppression or enhancement)
Instrument drift
Non-linearity at high concentrations
Incomplete recovery or adsorption
Method validation checks linearity, accuracy, precision, LOD/LOQ and matrix effects.
Key Takeaways
Quantitation converts ion intensity into amount or concentration.
Internal standards (especially isotope-labelled) are the most reliable approach.
Matrix effects and instrument condition are major sources of error.
Dynamic range and sensitivity set the practical limits.