Vacuum Physics

Physics

Vacuum Physics

What Vacuum Means in Mass Spectrometry

Vacuum is a space in which the pressure is significantly lower than atmospheric pressure. In a mass spectrometer the vacuum system removes most of the gas molecules so that ions can travel from the source to the detector with as few collisions as possible.

Pressure is commonly expressed in several units:

  • Atmosphere (atm) or bar ≈ 1000 mbar

  • Torr (1 Torr ≈ 1.33 mbar)

  • Pascal (Pa) or millibar (mbar)


Typical pressure regions relevant to mass spectrometry:

Region

Approximate pressure

Typical location in MS

Atmospheric

~1000 mbar

Outside the instrument, ESI source

Rough / low vacuum

1 – 10⁻³ mbar

Backing lines, some source regions

High vacuum

10⁻³ – 10⁻⁷ mbar

Most analyser regions

Ultra-high vacuum

< 10⁻⁷ mbar

Some high-performance analysers





Why High Vacuum Is Required

Ions must travel distances of centimetres to metres inside the instrument. At atmospheric pressure the average distance a molecule travels between collisions (the mean free path) is only tens of nanometres. Under those conditions an ion would collide thousands of times before reaching the detector and would be lost or scattered.

At high vacuum the mean free path becomes metres or longer, so most ions travel in straight lines without hitting residual gas molecules. This is essential for:

  • High ion transmission

  • Narrow peaks (good resolution)

  • Stable trajectories in analysers

  • Low chemical background


Residual Gas

Even in a good vacuum some gas molecules remain. Common residual gases include water, nitrogen, oxygen, hydrogen and hydrocarbons from outgassing of surfaces. These residual molecules can still cause occasional collisions, charge exchange or chemical noise. The composition of the residual gas matters as much as the total pressure.


Key Takeaways

  • Mass spectrometers require high vacuum so ions can travel without frequent collisions.

  • Analyser regions typically operate between 10⁻⁵ and 10⁻⁷ mbar.

  • Longer mean free path = fewer ion losses and better performance.

  • Residual gas is never zero; its nature and amount affect data quality.