top of page
d05cc009-8bfe-4bc2-94ff-a139a4a386c8.jpg

Learning Portal

BurgerMenuIcon copy.png
Massmend_Logo_Transparent.png

Mass Spectrometry Principles

Resolution

What Resolution Describes

Resolution (or resolving power) is the ability to separate ions that have very similar m/z values.


A common definition uses the peak width at half height (FWHM):


Resolving power = (m/z) / Δ(m/z)


Higher resolving power means narrower peaks and better separation of close masses.


Practical Meaning

  • Unit resolution separates peaks 1 Da apart.

  • High resolution (tens or hundreds of thousands) can separate ions that differ by only a few millidaltons.

  • Higher resolution improves confidence in elemental composition assignments and reduces interferences.


Trade-offs

Increasing resolution often reduces scan speed or sensitivity. The required resolution depends on the analytical goal.


Key Takeaways

  • Resolution measures how well close m/z values can be separated.

  • It is usually expressed as m/Δm.

  • Higher resolution helps with identification but may cost speed or sensitivity.